H2scan P&S Catalog 2025 - Flipbook - Page 9
HY-OPTIMA® 2700
EXPLOSION PROOF IN-LINE HYDROGEN PROCESS ANALYZER
Operating Conditions
Environmental
Ingress Protection
IP67
Operating Temperature
-20°C to 55°C (-4° F to 131°F)
Storage Temperature
-30°C to 80°C (-22° F to 176°F)
Humidity
0 to 95% RH, non-condensing
Mechanical
Dimensions H x W x D
190.5 x 137.06 x 144.78 mm (7.5 x 5.39 x 5.66 in)
Weight
2.7 kg (5.9 lbs)
Electrical
Power Input
90-240 VAC, 50-60 Hz
Input Power
15 W
Power Consumption
5 W (no external connection), up to 25 W (with 2 sensors connected)
Safety Certi昀椀cations
Analyzer
Remote Control
US/Canada
Class I, Div 1, Groups B, C, D
Class I, Div 1, Groups A, B, C, D
ATEX*
II 2 G Ex d IIB + H2 T4 Gb
II 1 G Ex ia IIC T4
IECEx*
Ex db IIB + H2 T4 Gb
Ex ia IIC T4 Ga
KOSAH
Ex db IIB + H2 T4
Ex ia IIC T4
UKCA*
II 2 G Ex d IIB + H2 T4 Gb
II 1 G Ex ia IIC T4
2575
0359
*Max pressure 1.1 ATM Absolute and max ambient temperature 55°C (131°F) at the analyzer
Performance
Operating Pressure at Analyzer
Recommended: 0.95-1.1 ATM Absolute (14.0-16.1 psia)
Maximum: 2 ATM Absolute (29.4 psia)
Process Gas Temperature
-20°C to 60°C (-4°F to 140°F)
Flow Rate
0.1 to 10 slpm
Operating Humidity
95% RH (non-condensing)
Calibration
90 days
Product Selection
Hydrogen Range
Accuracy
Drift/Week
Repeatability
Linearity
High
Hydrogen
MUST be
present
0.03%
10%
Yes