H2scan P&S Catalog 2025 - Flipbook - Page 13
HY-OPTIMA® 5030 Series
IN-LINE HYDROGEN PROCESS ANALYZER
Operating Conditions
Environmental
Ingress Protection
IP68
Operating Temperature (Ambient)
-20°C to 70°C (-4°F to 158°F)
Storage Temperature
-20°C to 105°C (-4°F to 221°F)
Mechanical
Dimensions H x W x D
149.8 x 40.64 x 40.64 mm (5.89 x 1.6 x 1.6 in)
Weight
0.4 kg (0.8 lbs)
Fittings
M12 4-wire. Fitting 3/4” - 14 MNPT
Electrical
Input Voltage
12-30 VDC
Input Power
10 W
Communication
Digital
RS485, 2-wire; 19200 baud, 8 bit data, 2 stop bits, no parity; Modbus protocol
Analog
4-20 mA available via optional accessory
Performance
Recommended Operating Pressure
0.95-1.1 ATM Absolute (14.0-16.1 psia)
Maximum Operating Pressure
10 ATM Absolute
Flow Rate
0.1 to 10 slpm (1/4” tube)
Operating Humidity
95% RH (non-condensing)
Calibration
Periodic calibration not required
Performance
CE, FCC, RoHS, WEE, Con昀氀ict Minerals
Accuracy/Repeatability Speci昀椀cations
Model
5031
Hydrogen Range
Repeatability
LDL-0.5%
0.05%
0.05%
0.5%-2%
0.1%
0.1%
2%-5%
0.15%
0.15%
5%-10%
0.2%
0.2%
Model
Hydrogen Range
5032
LDL-5%
Model
Hydrogen Range
5033
5034
Accuracy
Accuracy
0.3%
HY-OPTIMA 5030 Dimensions
Repeatability
0.3%
Accuracy
Repeatability
LDL-10%
0.4%
0.2%
10%-30%
0.7%
0.3%
30%-70%
1.0%
0.4%
70%-100%
1.2%
0.5%
Product Selection
Hydrogen Range
Model
Low
High
CO Limit
H2S Limit
T90
Response
Time (sec)
5031
0.03%
10%